Measuring Amplifier for pH and Redox type M3720

Product code: M3720
Category: pH, rH
QTY:
hotro

Minh Quan Technology Joint Stock Company

Hotline: 01692 114 683

Mr Que: 0973 852 363

Ms Ha: 0165 2197 433

Email: mqtekvietnam@gmail.com

Call us

Product information

  • 0…14pH, -1000…+1000mV
  • output 4…20mA, 2-wire-technics
  • 24VDC power supply
  • galvanically isolated
  • modbus
  • optional 2 – wire temperature compensation up to 100°
  • hold function
    Product Details

    The very small measuring amplifier for pH and Redox type M3720 converts the electrode signal to a 4…20mA 2-wire signal.

    The M3720 works with a resolution of 12 bit, is pre-calibrated and connected between the electrode and the signal acquisition instrument (e. g. PLC). Process control systems or any other control electronics may be connected to the amplifiers output.

    The electrode may be calibrated at the amplifier by a push button. Two LEDs indicate the calibration status. The hold function can freeze and release the output signal at any time.

    Sensor input and output are galvanically isolated from the output and power supply. The robust housing fulfills the protected class IP67 and allows application in rough environments.

    Optionally a 2-wire temperature compensation up to 100°C is available. pH/mV-electrodes with build-in temperature sensors are accepted by the amplifier.

    All settings and measured values may be read out and changed by the Modbus protocol.

Specifications

Measuring ranges: 0.00 to 14.

-1000 to +1000mV

Sensor inputs: Analog sensor, digital “Mettler Toledo” sensor on request
Accuracy: 0.1% @ 25°C
Stability: 0.003% per °C
Reproducibility: 0.1%
Linearity: 0.05%
Input impedance: 1000GΩ (1012 Ω)
Output rate: 10-90%: Typ. 10s, adjustable from 2s – 9min
Signal output:

–  Max load:

– Output impedance

4…20mA, adjustable over full range, others on request

500Ω @ 24V, 100Ω @ < 24V

Typ. >1MΩ

Calibration buffer solutions: Typical 4.00pH, 7.00pH, 10.00pH, others on request, Modbus programmable
Hold function: By double-clicking on the “Cal” button, the hold function for the current output can be turned on or off.
Temperature: 0…100°C without compensation or wire break: 25°C Automatically with an external transmitter. (PT100/PT1000/NTC30K) Types: NTC30k B = 3480, optional: NTC30k B = 4143
Operating temperature: -10°C … +40°C
Max. working temperature: -25°C … +65°C
Storage temperature -40°C … +85°C
Power supply range: 15 … 30VDC
Insulation resistance: > 5000MΩ @ 50VDC
Test isolation voltage 500VAC/1 minute (input -> output & power supply)
CE-Conformity: confirmed
EMV-Conformity: In accordance with EN61000-6-2, EN61000-6-4 und EN61326-1
Weight: ≈ 50g
Warranty 2 years
Material: anodized aluminum, IP67, others on request
Connection 2-wire signal: Cable ends, red = +24VDC, black = 4…20mA, shield
Connection Modbus: Cable ends, brown = RS485A, orange = RS485B
Connection electrode: Coax core = pH, Coax shield = reference
Cable length 2-wire signal: 300m, 3m factory-made
Cable length electrode: 60cm, factory-made, connected
Options: Customer specified functions/cable connection

Modbus

Connection for Mettler Toledo digital sensors

Ordering example: 1x M3720-pH 0…14pH = 4…20mA

1x M3720-mV -1000mV…+1000mV = 4…20mA

1x M3720-Modbus

 