Slide ảnh 1
Slide ảnh 2

Product

Conductivity meter with USB logger type M3036

Product code: M3036
Category: Conductivity
QTY:
hotro

Minh Quan Technology Joint Stock Company

Hotline: 01692 114 683

Mr Que: 0973 852 363

Ms Ha: 0165 2197 433

Email: mqtekvietnam@gmail.com

Call us

Product information

The M3036 conductivity meter is suitable for water, waste water or pure water conditioning in continuous or batch-type operating modes, for liquid chromatography or for general chemical process monitoring.

  • Conductivity meter for 0 … 20/200/2000/20000/200000uS
  • USB data logger
  • Galvanically separated sensor cirquit
  • Galvanically separated signal output. free configurable between 0 and 20mA.
  • Precision 3-wire PT100 temperature input
  • Up to 5 limit contacts
  • Universal power supply: All voltages between 20VAC/DC and 253VAC/DC possible

Specifications

Measuring ranges: 0…2.000µS C = 0.01, C = 0.1

0…20.00µS C = 10.0, C = 1.0, C = 0.1, C = 0.01

0…200.00µS C = 10.0, C = 1.00, C = 0.1

0…2.000mS C = 10.0, C = 1.00, C = 0.1

0…20.00mS C = 10.00, C= 1.0

0…200.0mS C = 10.0

Accuracy: 0.50%
Display: Graphic LCD 128×128 pixel
Pixel size: 0.4×0.4mm
Temperature drift: Max. 50µV/°C
Max. humidity: 95%, non condensing
Temperature compensation: Manual from 0 to 130°C.

Automatic with external Pt-100, 3-wire sensor. Wire break of the Pt

100 sensor switches the device to 25°C. Manual / automatic with (F4)

Temperature slope: 0.00%/°C (no compensation) to 8.00%/°C
Self conductivity of water: controlled and compensated to 25°C
Reference temperature 25°C
Max. length of cell cable: Cable capacity is compensated until max capacity of 10nF or (10-8F)
ENT key: Quick access to settings or to save parameters
ESC key: Access to the main menu, press and hold for 3 seconds, or exit menu item
+ / – key: Set values
F3 key: Change measuring range of measuring value, limit contacts and alarm limits
Programming functions: Company label and various setting options via software
Signal current output:

– Max load:

– Output impedance:

0…20mA programmable, isolated against electronic ground

<500Ω

Typ. >1MΩ

Relay contacts: Adjustable over the full range
Hysteresis: 5 – 200 digits adjustable
Limit value status Displayed with green LEDs for contact 1/2, no LEDs for optional contact 3/4
Relay outputs: Floating changeover contacts, max. 6A, 2A at 230VAC continuous inductive load
Alarm functions: 2 programmable limits, Pt-100 alarm if wires are broken, ‘1Wire’ alarm if wires are broken, alarm if the pump’s max. run time times out and an USB alarm if the stick/media is full. The alarm is displayed with the red LED and on display. Alarm can be acknowledged.

 