Temperature
Thermo couple transmitter type M8841
Minh Quan Technology Joint Stock Company
Hotline: 01692 114 683
Mr Que: 0973 852 363
Ms Ha: 0165 2197 433
Email: mqtekvietnam@gmail.com
Call usProduct information
- cold junction compensation
- sensor operating range: -100…1800ºC
Specifications
Linearization: | 13-bit, digital | |
Linearization error: | typ. 0.1% | |
Input voltage: | Maximum ±100 mV | |
Input protection: | Diode protection | |
Input impedance: | 1GΩ | |
Damping: | Maximum bandwidth 0.2 Hz | |
Accuracy: | ±1°C or ±12 mV whichever is greater | |
Zero error: | Maximum ±5 mV/°C (chopper amplifier | |
Gain error: | typ. 50 ppM/°C | |
Cold junction compensation: | 1. No compensation. Wire Jumper “X” from 4 to 5.
2. Internal compensation with Pt-1000 platinum probe -10 to +50°C (14 to 122°F). Wire Jumper “Y” from 3 to 6. 3. External compensation with Pt-1000 platinum probe. The measurement is made in the thermocouple junction box. Compensated junction box temp. range -10 to +50°C (14 to 122°F). Connect the external Pt-1000 platinum probe to 4 and 6. |
|
Signal current output : | 0…20mA / 4…20mA, others on request | |
Current outout impedance | >1MΩ at 20mA | |
Current output load: | >750Ω | |
Supply voltage: | 115V/60 Hz or 230V/50 Hz, ca. 5 VA, others on request | |
Terminals: | 12-pole screw terminal | |
Terminal description: | 1 = Signal Input +
2 = Signal Input – 3 = Cold junction comp. A 4 = Cold junction comp. B 5 = Cold junction comp. C 6 = Cold junction comp. D |
7 = Signal output +
8 = Signal output – 9 = not used 10 = not used 11 = Power supply N 12 = Power supply |
Mounting | 35mm mounting rail, EN50022-35 | |
Weight: | 280 grams (9.8 oz) | |
Warranty: | 2 years | |
Options: | – Special supply voltage
– Special ranges for thermocouples below – Special Output Signal currents – Special Input Signals supplied by customer – Pt-1000 platinum probes for junction boxes |
Reviews
Related Products
Load cell type M2428-14
Code: Product Details The M2428-14 is a low-cost load cell to measure weights or forces in the range of 25kg (~250N) to 30'000kg (~300'000N). The bending beam consists of a strain gage sensor with integrated electronics to convert the force or weight to a 4…20mA signal. There is no need to use external, sensitive strain gage amplifiers or signal conditioning devices. A cell connects directly to any 2-wire sensor input of process control computers or other control equipment and displays. The bending beam confirms the protected class IP67 an delivered with open cable ends. Any range within the measuring range can be programmed. E.g. 45kg to 420kg = 4 to 20mA within a calibrated range of 0 to 500kg.